Ametek Crystal HPC51 Operation Manual page 26

Intrinsically safe handheld pressure calibrator
Table of Contents

Advertisement

Differential Pressure
X
The Tare function can improve differential pressure measurement uncertainties. Requires the use of an equalizing valve. Use the higher range sensor specifica-
tion when two different ranges are selected.
Full Scale Range of Both Sensors
(psi)
15
0.00015
30
0.0005
100
0.0015
300
0.005
1000
0.02
3000
0.05
10000
Without tare function:
± (0 . 0 5% of static line pressure reading)
Barometric Reference (BARO)
Accuracy . . . . . . . . . . . . . . . . . . . . . . . . ± 0.00725 psi, ± 0.5 mbar
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.153 to 15.954 psiA, 700.0 to 1100.0 mbarA
Units and Resolution . . . . . . . . . . . . . psi:
inHg: 0.001
mmHg: 0.01
mbar
Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range for
one year.
Note:
Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.
Pressure Sensor (P1, P2, and APMi)
Wetted Materials. . . . . . . . . . . . . . . . . (Wrench Tight) 316 stainless steel
(Finger Tight) 316 stainless steel and Viton® (internal o-ring)
(15 psi /1 bar /100 kpa) 316 stainless steel and Viton® (internal o-ring)
Diaphragm Seal Fluid: . . . . . . . . . . . . Silicone Oil
Connection: . . . . . . . . . . . . . . . . . . . . . Crystal CPF Female
The Greater of (+/–)
psi
mbar
inH
O
mmH
2
0.01
0.004
0.04
0.014
0.10
0.04
0.4
0.14
1.0
0.4
10.0
4.0
1.4
0.2
10.0
4.0
Unit is enabled in CrystalControl
0.001
0.1
O
% of DP Reading
2
0.1
0.4
1.0
4.0
or
0.035%
n/a
n/a
24
Specifications
HPC50 Series Operation Manual