Avance
Message
Priority
Display panel
Medium
controls failure
EtCO2 high
High
EtCO2 low
Informational
Medium after 1
minute
EtDES high
Medium
EtDES low
Informational
EtENF high
Medium
EtENF low
Informational
EtHAL high
Medium
EtHAL low
Informational
EtISO high
Medium
EtISO low
Informational
EtO2 high
Medium
EtO2 low
Medium
EtSEV high
Medium
EtSEV low
Informational
FiCO2 high.
High
Absorbent OK?
FiDES high
Medium
7-6
Cause
Communication lost between
panel and key pad.
EtCO2 is greater than high alarm
limit.
EtCO2 is less than alarm limit.
Priority escalates to Medium if
alarm is not resolved after 1
minute.
EtDES is greater than alarm
1
limit.
EtDES is less than alarm limit.
EtENF is greater than alarm
1
limit.
EtENF is less than alarm limit.
EtHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
EtHAL is less than alarm limit.
EtISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
EtISO is less than alarm limit.
EtO2 is greater than high alarm
limit.
EtO2 is less than low alarm limit. Set the alarm limits appropriately. Increase
EtSEV is greater than alarm
limit.
EtSEV is less than alarm limit.
FiCO2 is greater than alarm limit. Check if absorbent needs to be changed.
FiDES is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
Action
Turn the system off and back on.
Check the patient and EtCO2 settings. Check
if absorbent needs to be changed.
Ensure the patient is properly intubated.
Check for leaks or blockages in the patient
circuit.
Set the alarm limits appropriately. Decrease
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
Set the alarm limits appropriately. Decrease
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
Set the alarm limits appropriately. Decrease
the O
concentration.
2
the O
concentration.
2
Set the alarm limits appropriately. Decrease
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
Check the patient.
the agent concentration.