Datex-Ohmeda Avance User's Reference Manual page 101

Hide thumbs Also See for Avance:
Table of Contents

Advertisement

Message
Priority
FiDES low
Informational
FiENF high
Medium
FiENF low
Informational
FiHAL high
Medium
FiHAL low
Informational
FiISO high
Medium
FiISO low
Informational
FiO2 high
Medium
FiO2 low
High
FiSEV high
Medium
FiSEV low
Informational
For mech vent,
Informational
set Bag/Vent
switch
Gas monitoring
Medium
not available
Increase low MV
Medium
limit
Inspiration
Medium
stopped
Internal failure.
High
System may
shut down.
Internal failure.
Medium
System may
shut down.
M1145956
Cause
FiDES is less than alarm limit.
FiENF is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
FiENF is less than alarm limit.
FiHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
FiHAL is less than alarm limit.
FiISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
1
FiISO is less than alarm limit.
FiO2 is greater than high alarm
limit.
FiO2 is less than low alarm limit. Check O
FiSEV is greater than alarm limit. Set the alarm limits appropriately. Decrease
FiSEV is less than alarm limit.
Bag/Vent switch is set to Bag
and the ACGO switch is set to
Circle.
Airway module hardware failure. Replace airway module. Then, turn power off
Low MV limit is off in SIMV/PSV,
SIMV-PC, or PSVPro modes.
High airway pressure.
Power controller software failure. Contact a Datex-Ohmeda trained service
Power controller software failure. Contact a Datex-Ohmeda trained service
7 Alarms and Troubleshooting
Action
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
Check the O
setting. Recalibrate the O
2
and the airway module.
setting. Check for leaks or
2
blockages in the patient circuit.
the agent concentration.
Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
Move switch to the Vent position to start
mechanical ventilation.
and back on between cases to clear the
alarm and receive module data.
Increase Low MV alarm limit to improve
patient disconnection detection.
Check system for blockages.
representative.
representative.
cell
2
7-7