M1145956
9. Test the Ppeak low. Leak? alarm:
Unblock the patient connection.
Set the Bag/Vent switch to Vent.
Set the tidal volume and total flow to minimum.
Other alarms such as MVexp low can occur.
Make sure that the Ppeak low. Leak? alarm occurs.
10. Set all alarm limits to approved clinical values.
7 Alarms and Troubleshooting
7-15