Default settings for default case types
Name
Patient and Sensor Type
Patient Weight
Patient Age
Other Gas
Circuit
Data Source
Vent Mode
TV
RR
I:E
Tpause
PEEP
Pmax
Pinsp
RR
I:E
PEEP
Pmax
Rise Rate
O2 %
Total Flow
Sweep Speed
Split Screen
Waveform Field 1
Waveform Field 2
Waveform Field 3
Digit Field
Auto MV Limit
Alarm Volume
MV/TV Alarms
Volume Apnea
MV High
MV Low
TV High
TV Low
RR High
RR Low
EtCO2 High
EtCO2 Low
FiCO2 High
FiO2 High
FiO2 Low
EtO2 High
EtO2 Low
FiISO High
M1145956
Each case type has multiple settings. The default settings for the
default case types show in the following table. Values in bold are
different from the default case type ADULT. A * indicates that the
setting is not available for the default ventilation mode.
ADULT
PEDIATRIC
ADULT
PEDIATRIC
Adult
Pedi
70 kg
18 kg
40 y
5 y
Air
Air
Circle
Circle
Vent
Vent
VCV
PCV
500
*
12
*
1:2
*
Off
*
Off
*
40
*
*
5
*
12
*
1:2
*
Off
*
40
*
Auto
100
100
6.00
6.00
Fast
Fast
Paw
Gas
Paw
Paw
Flow
Flow
CO2
CO2
AA
AA
Off
Off
3
3
On
On
On
On
10.0
10.0
2.0
2.0
1000
1000
Off
Off
Off
Off
Off
Off
8.0
8.0
3.0
3.0
Off
Off
Off
Off
21
21
Off
Off
Off
Off
5
5
12 Super User Mode
LOCAL
CUSTOM 1
LOCAL
CUSTOM 1
Adult
Adult
70 kg
70 kg
40 y
40 y
Air
Air
Circle
Circle
Vent
Vent
VCV
VCV
500
500
12
12
1:2
1:2
Off
Off
Off
Off
40
40
*
*
*
*
*
*
*
*
*
*
*
*
100
100
0.20
6.00
Fast
Fast
Gas
Gas
Paw
Paw
Flow
Flow
CO2
CO2
AA
AA
Off
Off
1
3
Off
On
Off
On
Off
10.0
Off
2.0
Off
1000
Off
Off
Off
Off
Off
Off
8.0
8.0
3.0
3.0
Off
Off
Off
Off
21
21
Off
Off
Off
Off
5
5
12-13