SIMV-PC mode
M1145956
11 Specifications and Theory of Operation
Synchronized Intermittent Mandatory Ventilation, Pressure Control is
a mode in which a relatively slow mandatory breathing rate is set with
pressure-controlled breathing. This mode combines mandatory
breaths with spontaneous breath support. If a trigger event occurs
within the synchronization window, a new pressure-controlled breath
is initiated. If a trigger event occurs elsewhere during the expiratory
phase, a support for a spontaneous breath is provided with pressure
support added as set by the clinician.
SIMV-PC mode settings:
•
Pinsp
•
RR
•
Tinsp
•
Psupport
•
PEEP
•
Pmax
•
Trig Window
•
Flow Trigger
•
End of Breath
•
Rise Rate
•
Exit Backup
1. Paw waveform
2. Tinsp
3. Spontaneous breathing period
4. Trigger window
5. Pressure supported breath
6. Pinsp
7. Flow waveform
Figure 11-5 • SIMV-PC waveforms
11-23