Message
Priority
Calibrate
Informational
(remove) flow
sensors
Calibrate O2
Informational
sensor
Calibrate, dry, or
Informational
replace flow
sensors
Cannot monitor
Medium
Air pipeline
Cannot monitor
Informational
gas supplies
Cannot monitor
Medium
O2 pipeline
Cannot read gas
Medium
supply
pressures
Check circuit
Medium
connections
Check D-Fend
Medium
Check flow
Medium
sensors
Check sample
Medium
gas out
Circuit leak
Medium
Circuit leak
Informational
silenced
Circuitry >75C
Medium
shutdown
possible
Cooling fan
Medium
needs service.
System OK.
Cooling fans
Medium
failed. May
overheat.
M1145956
Cause
Flow calibration failure or more
than 24 hours have passed since
the last flow sensor calibration.
Calibration failure or measured
O2 is greater than 110%.
Patient volume mismatch
occurred during the last case.
Air pipeline pressure is invalid.
Hardware failure.
O
pipeline pressure is invalid.
2
Transducer failure.
Breaths detected in circle circuit
while non-circle circuit is
selected.
Water trap not attached.
System has detected an
improper flow pattern in the
breathing circuit.
Possible blockage in airway
module sample gas outlet.
Vent TVexp is less than half of
vent TVinsp for at least 30
seconds.
Setting on Alarm Setup menu.
Power supply temperature
exceeds 75°C.
Fan reporting error.
Fan reporting error.
7 Alarms and Troubleshooting
Action
Calibrate the flow sensors.
Calibrate the O2 cell. Replace the O2 cell if
necessary.
Calibrate, dry or replace the flow sensors.
Start a new case.
Check pipeline supply pressure.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
Check pipeline supply pressure.
Contact a Datex-Ohmeda trained service
representative.
Check circle circuit connections and settings.
Check that the water trap is properly attached
to the airway module.
Ensure the internal flow sensors are
connected correctly.
Check for blockage in the airway module
sample gas outlet. Remove blockage.
Check for leaks in the patient circuit.
Calibrate flow sensors. If problem persists,
replace flow sensors.
Vent TVexp is less than 50% of vent TVinsp
for at least 30 seconds. Message indicates
that the Leak Audio alarm is turned off.
Shut down system as soon as possible.
Then, check cooling fans and filters.
Shut down system as soon as possible.
Then, check cooling fans and filters.
Shut down system as soon as possible.
Contact a Datex-Ohmeda trained service
representative.
7-5