EZchange canister (optional)

M1145956
Push the absorber canister release to activate the EZchange canister
mode. The canister will swing down to the EZchange position. The
EZchange canister mode seals the breathing circuit when the canister
holder is down. This permits continued ventilation and rebreathing of
exhaled gases while easily replacing the absorber canister.
Systems with EZchange canister have the following label on the
canister holder. When the system is in EZchange position, the
message 'CO2 Absorber Out of Circuit' shows in the waveform area
on the anesthesia display.
To return to absorber mode, reinsert the canister into the holder and
push the canister back up and snap it into absorber position. When
the canister is in the absorber position, the exhaled gas flows through
the absorber, removing CO2.
Note
Check the absorber canister to ensure it has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
1. Side rails
3 Operation
1
3-27