iCAP 7000 Series ICP-OES Spectrometer
Open the Sprint Valve Configurator
software from the desktop icon.
Click the "connect to sprint valve"
button
Set the following parameters:
Use a 4ml sampling loop
Loop Evacuation Delay = 2.0
Loop Load = 8.0
Equalization Delay = 1.5
Probe Rinse = 1.5
Rinse Station Refill = 10
Set the autosampler uptake time = 20
(To allow the valve to fill and the sample
to reach the plasma, this is set in
Qtegra)
Press "save configuration to sprint
valve"
If most of the options are grayed out click "Connect to Sprint Valve".
Expand the window using the arrow
button on the right side of the window to
open manual controls.
Page 32 of 36
Installation Details