Resetting The Password Or Device Via Operating Keys; Commissioning With "Commissioning" Wizard; Storing The Oscillation Frequencies; Simulation - Endress+Hauser Liquiphant FTL51B Operating Instructions Manual

Vibronic hart point level switch for liquids
Hide thumbs Also See for Liquiphant FTL51B:
Table of Contents

Advertisement

Commissioning
1)
DeviceCare is available for download at www.software-products.endress.com. You must register in the Endress+Hauser software portal to
download the product.
34
9.7.2

Resetting the password or device via operating keys

Resetting the password
1. Briefly press operating key I 3 times within 4 seconds.
 The (green) LED flashes twice repeatedly at short intervals.
2. Press operating key I again to confirm and release the key.
 The password has been reset.
The LED stops flashing.
 If the time until confirmation is longer than 15 s, the device quits the "Reset Password"
state and the LED stops flashing. The password is not reset. Repeat procedure if necessary.
Resetting the device (order configuration)
Press both operating keys simultaneously for at least 12 s.
 The device has been reset (as-delivered state).
9.7.3

Commissioning with "Commissioning" wizard

1)
In FieldCare, DeviceCare
available to guide the user through the initial commissioning steps.
1. Connect the device to the operating tool.
2. Open the device in the operating tool.
 The dashboard (homepage) of the device is displayed:
3. Open the wizard: Guidance menu → Commissioning wizard
4. Enter the appropriate value in each parameter or select the appropriate option. These
values are written directly to the device.
5. Click "Next" to go to the next page.
6. Once all the pages have been completed, click "End": the Commissioning wizard is
closed.
If the Commissioning wizard is cancelled before all the necessary parameters have
been configured, the device may be in an undefined state. In such situations, it is
advisable to reset the device to the factory default settings.
9.7.4

Storing the oscillation frequencies

Two frequencies (uncovered/covered) can be stored in the device so that the current
oscillation frequency can be compared later on to the condition at the time of
commissioning.
The frequencies can only be stored in the respective fork state. For example, if the fork is
covered it is only possible to save the frequency when the fork is covered (Stored covered
frequency parameter).
The value is saved via the Commissioning wizard or via
Navigation: Application → Sensor → Stored frequency
9.8

Simulation

In the Simulation submenu the user can simulate with the options Fork state (uncovered/
covered), Sensor frequency, Current output and Diagnostic event simulation.
, SmartBlue and on the display, the Commissioning wizard is
Liquiphant FTL51B HART
Endress+Hauser