Liquiphant FTL51B HART
Endress+Hauser
Initial password
The serial number of the device serves as the initial password when the connection is
established for the first time.
Please note the following
If the Bluetooth display is removed from one device and installed in another device:
• All the log-in data are only saved in the Bluetooth display and not in the device
• The password changed by the user is also saved in the Bluetooth display
7.6
Access to the operating menu via the operating tool
Access via the operating tool is possible:
• Via HART communication, e.g. Commubox FXA195
• Via Endress+Hauser Commubox FXA291
With the Commubox FXA291, a CDI connection can be established with the device
interface and a Windows PC/notebook with a USB port
7.7
HistoROM data management
When replacing the electronic insert, the stored data is transferred by reconnecting the
HistoROM.
The device serial number is saved in the HistoROM. The electronics serial number is saved
in the electronics.
8
System integration
8.1
Overview of device description files
• Manufacturer ID: 17 (0x11)
• Device type code: 0x11C4
• Device revision: 1
• HART specification: 7
• DD revision: 1
• Device description files (DTM, DD) information and files at:
•
www.endress.com
•
www.fieldcommgroup.org
• HART load: min. 250 Ω
8.2
Measured variables via HART protocol
The following measured values are assigned to the device variables at the factory:
Primary variable (PV) parameter (Primary variable)
Secondary variable (SV) parameter (Secondary variable)
Device variable
System integration
Measured value
1)
Level limit detection
Sensor frequency
2)
3)
29