Commissioning
32
9.4.2

FieldCare/DeviceCare via service interface (CDI)

1
1
Computer with FieldCare/DeviceCare operating tool
2
Commubox FXA291
3
Service interface (CDI) of the device (= Endress+Hauser Common Data Interface)
At least 22 mA is required to update (flash) the device firmware.
9.5

Configuring the device address via software

See HART address parameter.
Navigation: Application → HART output → Configuration → HART address
9.6

Setting the operating language

9.6.1

Local display

Setting the operating language
To set the operating language, the display first needs to be unlocked:
1. Press the  key for at least 2 s.
 A dialog box appears.
2. Unlock the display operation.
3. Select the Language parameter in the main menu.
4. Press the  key.
5. Select the desired language with the  or  key.
6. Press the  key.
9.6.2

FieldCare

1. In the "Extras" menu, click "Options".
2. Set the preferred language for FieldCare in the "Language" section.
Setting the language for the local display via FieldCare
Navigation: System → Display → Language
Set the desired language in the Language parameter.
9.6.3

DeviceCare

Click the menu icon:
Liquiphant FTL51B HART
2
3
A0039148
Endress+Hauser