Error E32 displayed on DX-D Software Console after
preparing for exposure
Details
Cause
Brief Solution
Error E32 displayed on DX-D Software Console after
preparing for exposure.
The bucky of the X-ray system is not closed.
The bucky of the X-ray system is empty.
The CR cassette has not been replaced after the expo-
sure.
Check if the bucky of the X-ray system contains an un-
exposed CR cassette or a DR detector and if it is proper-
ly closed.
Click the OK button on the DX-D Software Console or
the Reset button on the DX-D 400 Touchscreen Console
or on the DX-D 400 overlay console until the error indi-
cation disappears on the console.
If the equipment remains inoperative, try restarting the
generator or call service. In this case, the generator can
still work in free exposure mode.
DX-D 400 | Problem solving | 71