AGFA DX-D 400 User Manual page 45

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

Warning: Avoid unnecessary dose by checking the
workstation selection on the X-ray generator console before
exposing. In a configuration with a DR Detector configured
on a virtual port, the DR Detector will not be triggered if a
free exposure is selected on the Generator console and yet
the exposure will be allowed.
Warning: As the cables of the equipment are long, be careful
not to entangle the cables during use. Also, be careful not to
trip over the cables. Falls could result in injury.
Warning: Unplugging the detector immediately after
exposure may cause image loss.
Warning: To prevent unintended exposure, the position of
the exposure footswitch should be such that it cannot be
accidentally stepped on.
Caution: When operating the DR detector, the calculated
exposure time (ms) or manual overrides should never exceed
the maximum exposure time (Max ms) specified as integration
time of the DR detector.
Caution: Make sure the patient hand grips are properly
mounted.
Caution: Excessive ambient temperature may impact
performance of DR Detectors and cause permanent damage to
the equipment. If ambient temperature and humidity is outside
the range specified in the technical data, do not operate the
system or use air conditioning. Warranty will be void if it is
obvious that operating conditions are not met.
Caution: Damaged grid. Reduced image quality. Please handle
the grids with special care.
Caution: When inserting the scattered radiation grids, it is
essential that the grid corresponds to the intended source-
image-distance (SID) to which the grid is focussed. Because of
the focussing of the grids, the tube unit must be centered onto
the bucky.
DX-D 400 | Introduction to DX-D 400 | 45