AGFA DX-D 400 User Manual page 81

Hide thumbs Also See for DX-D 400:
Table of Contents

Advertisement

Note: In accordance with Standard IEC 61601-1-2, the emissions character-
istics of this equipment make it suitable for use in industrial areas and hospi-
tals (CISPR 11 Class A. If it is used in a residential environment (for which
CISPR 11 Class B is normally required) this equipment might not offer ade-
quate protection to radio-frequency communication services. The user might
need to take mitigation measures, such as relocating or re-orientating the
equipment.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such an
environment.
Table 3: Guidance and manufacturer's declaration on electromagnetic
immunity
Immunity Test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
IEC 60601-1-2
Compli-
Test Level
± 6 kV contact
± 6 kV
discharge
± 8 kV
± 8 kV air dis-
charge
± 2 kV for pow-
± 2 kV
er supply lines
N/A
± 1 kV for in-
put/output
lines
± 1 kV line(s)
± 1 kV
to line(s)
± 2 kV
± 2 kV line(s)
to earth
40% U
(60%
60% for
T
dip in U
) for 5
5 peri-
T
cycles
ods
70% U
(30%
30% for
T
dip in U
) for
25 peri-
T
25 cycles
ods
DX-D 400 | Technical Documentation | 81
Electromagnetic environ-
ance
ment - guidance
Level
Floors should consist of
wood, concrete or ceramic
tiles. The relative humidity
must be at least 30%, if the
floor is made of synthetic
material.
The quality of the voltage
supplied should corre-
spond to a typical commer-
cial or clinical environ-
ment.
The quality of the voltage
supplied should corre-
spond to that of a typical
commercial or clinical envi-
ronment.
The quality of the voltage
supply should correspond
to that of a typical commer-
cial or clinical environ-
ment. If the user wants the
device to work continuous-
0230D EN 20191209 1135