NX receives black or underexposed DR image due to
repeatedly pushing the exposure button
Details
Cause
Brief Solution
Related Links
Performing an exposure using the DR Detector
A black or underexposed image is arriving on the NX
workstation.
The exposure button was pushed to the first pressure
point and released without making an exposure.
Directly afterwards, the exposure button was pushed
down fully.
The X-ray system may need a longer preparation time
directly after an interrupted preparation cycle. This
prevents the DR Detector to synchronize with the X-ray
system.
Depending on the X-ray system, two situations can oc-
cur:
• The X-ray system will not make the exposure and
the DR Detector acquires an image without expo-
sure.
• The X-ray system will start the exposure with a de-
lay and the DR Detector cannot acquire the com-
plete dose.
To repeat the exposure workflow:
1. On the NX workstation, click Copy Exposure to
create a new thumbnail.
2. Repeat the steps described in the Basic Workflow.