Wrong modality position selected

Details
Cause
Brief Solution
The active modality position on the X-ray system does
not match the selected modality position on the NX
Workstation.
The modality position has been modified on the gener-
ator console.
This situation applies only to Siemens generators.
To use another modality position for a scheduled expo-
sure:
1. On the NX workstation, click Edit in the Image De-
tail Pane and select an exposure type for the correct
modality position.
2. Continue the exposure workflow.
DX-D 400 | Problem solving | 69
0230D EN 20191209 1135