Figure 15: Portrait orientation in the radiographic wall stand right loading

Orientation of DX-D 10C, DX-D 10G in the bucky

To avoid damage to the cable of the detector, there are restrictions on the
orientation of the detector when loading the bucky.
Topics:
Orientation in the radiographic table
Orientation in the radiographic wall stand left loading
Orientation in the radiographic wall stand right loading
Orientation in the radiographic table
To use the detector in landscape orientation, insert the detector in landscape
orientation with the cable at lower right hand side.
To use the detector in portrait orientation:
1. Insert the detector in landscape orientation with the cable at lower right
hand side.
2. Rotate the detector in the bucky.
1
Figure 16: Portrait orientation in the radiographic table
1
CAUTION:
Inserting DX-D 10C, DX-D 10G using other orientations than
described will damage the cable when closing the bucky or
when rotating the carrier.
2
2
3253F EN 20210312 1125
Bucky | Bucky | 39