Appendix A: Service Guide - System Evaluation

Lot/Serial #
Controller Evaluation:
Controller Serial #
Step
Criteria
1
The Controller performs as per instructions in the User Manual
2
Light Module Use is <24,900 hrs.
Record Light Module
Hours:
Light Bed Evaluation:
Light Bed Serial #
Step
Criteria
1
GelMat is clear and not yellow/discolored
2
GelMat is adhered to the Light Bed and not damaged (tears or holes)
3
GelMat is no older than 6 months, has not undergone more than 1,500
treatment hours, or does not affect calibrated light intensity levels.
4
All the measured, averaged, and adjusted light intensities per section 4.3
pass the acceptance criteria
Service Performed/Notes:
© 2022 by NeoLight LLC.
All rights reserved.
LM1: _________(hrs.)
Date
LM2: __________(hrs.)
Pass
Fail
Pass
Fail
Page 42 of 43
500050 Rev A