NeoLight Skylife Service Manual page 32

Infant phototherapy system
Hide thumbs Also See for Skylife:
Table of Contents

Advertisement

System Indicators
Functionality Issues
- Warning Message:
"!! SYSTEM FAILURE
00002"
- Warning Message:
"!! SYSTEM FAILURE
00004"
© 2022 by NeoLight LLC.
All rights reserved.
Likely Cause
LED Bank 1 Current
Line Failure
LED Bank 2 Current
Line Failure
Trouble Shooting Steps
1.
Follow procedures to open and access the Light Modules in
the Replacement instructions in Chapter 4.5. Then,
carefully detach Data Port 1 per instructions.
2.
Data Port 1 is designed with a jumper wire between pin 8
and 7 on the cable connector
3.
Inspect the connector for sufficient contact.
4.
If the connector is not seated correctly, reconnect the bank
properly per the provided instructions in Chapter 4.5 and
restart the device
5.
If wiring damage is seen or if the error persists after
correction, contact NeoLight
1.
Refer to instructions provided in error 00002 but locate
Data Port 2.
2.
Data Port 2 is designed with a jumper wire between pin 6
and 7 on the cable connector
Page 32 of 43