Performance Verification - NeoLight Skylife Service Manual

Infant phototherapy system
Hide thumbs Also See for Skylife:
Table of Contents

Advertisement

4.3 Performance Verification

IMPORTANT!
The light output of Skylife was factory calibrated with the Ohmeda Medical BiliBlanket
Light Meter ii. Because your facility may use a different Spectroradiometer to measure
light intensity output, it is necessary to understand how your reading correlates to the
Ohmeda Medical BiliBlanket Light Meter ii reading. The following table provides a light
intensity output conversion chart.
GE Ohmeda BiliBlanket Light
Meter ii
(μW/cm
2
/nm)
20.0
25.0
30.0
35.0
40.0
45.0
50.0
55.0
60.0
65.0
70.0
75.0
80.0
I
GE
IMPORTANT!
Skylife is factory calibrated with the disposables in place. Therefore, light intensity
testing must be performed with a Skylife Disposable Swaddle.
Please ensure Skylife is operating at optimum levels using the following procedure:
1. Skylife's performance is evaluated by measuring its light intensity across twenty-six 1x1" cells at each
treatment setting (Low, High, and Very High)
© 2022 by NeoLight LLC.
All rights reserved.
Natus® neoBLUE® Radiometer
(μW/cm
2
/nm)
18.1
22.4
26.8
31.2
35.5
39.9
44.2
48.6
52.9
57.3
61.7
66.0
70.4
Conversion Formula:
I
I
= 0.8713*
+ 0.6617
NN
GE
Respironics Joey Dosimeter
(μW/cm
2
/nm)
52.7
65.8
78.9
92.0
105.1
118.2
131.3
144.4
157.5
170.6
183.7
196.9
210.0
Conversion Formula:
I
I
= 2.6214 *
+ 0.2450
RJ
GE
Page 22 of 43
500050 Rev A