System Indicators
Functionality Issues
- Warning Message:
"!! SYSTEM FAILURE
00008"
- Warning Message:
"!! SYSTEM FAILURE
00016"
- Warning Message:
"!! SYSTEM FAILURE
00024"
- Warning Message:
"!! SYSTEM FAILURE
00032"
© 2022 by NeoLight LLC.
All rights reserved.
Likely Cause
Bank 1 temperature
1.
sensor error
2.
3.
Bank 2 temperature
sensor error
Bank 1 and Bank 2
1.
Temperature Sensor
Error
Bank 1 EEPROM error
1.
2.
Trouble Shooting Steps
Follow procedures in Chapter 4.5 to open and access the
Light Modules and follow the procedure to remove Light
Bank 1.
If the thermostat module is disconnected, reconnect the
thermostat and Light Bank, and restart the system.
Important! Do not press on the wires down during this
process.
If the thermostat module was not disconnected, try
disconnecting the module using either tweezers or a small
flat head screwdriver. Then, reconnect and restart the
system.
If the wires are damaged or if the issue persists after
reconnection, contact NeoLight
1.
Refer to the instructions provided for SYSTEM FAILURE
00008 but access Light Bank 2
See instructions on how to troubleshoot SYSTEM FAILURE
00004 and SYSTEM FAILURE 00008
Restart the system
If the error persists, contact NeoLight