NeoLight Skylife Service Manual page 34

Infant phototherapy system
Hide thumbs Also See for Skylife:
Table of Contents

Advertisement

System Indicators
Functionality Issues
- Warning Message:
"!! SYSTEM FAILURE
00064"
- Warning Message:
"!! SYSTEM FAILURE
00120"
- Warning Message:
"!! SYSTEM FAILURE
00128"
- Warning Message:
"!! SYSTEM FAILURE
00256"
Any other error not
specified above
© 2022 by NeoLight LLC.
All rights reserved.
Likely Cause
Bank 2 EEPROM error
1.
2.
Loose Controller J1
1.
Connector
Stalled Fan
Temperature sensor
error
1.
Trouble Shooting Steps
Restart the system
If the error persists, contact NeoLight
Contact NeoLight
1.
Restart the system
2.
If the error persists, contact NeoLight
1.
Turn off the device
2.
Check if vents are blocked.
3.
Clear the area around the device and restart the device
after 10 mins.
4.
If the issue persists, contact NeoLight.
Contact NeoLight
Page 34 of 43
500050 Rev A