Skylife™ is intended for use in the electromagnetic environment specified below. Therefore, the customer or
the user of Skylife™ should ensure that it is used in such an environment.
Emission Tests
RF emissions CISPR 11
R.F. emissions CISPR 11
Harmonic emissions IEC 61000-3-
2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Skylife™ is intended for use in the electromagnetic environment specified below. Therefore, the customer or
the user of Skylife™ should ensure that it is used in such an environment.
Immunity Test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient
/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC
61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
NOTE: U
is the A/C mains voltage prior to application of the test level.
T
Skylife™ is intended for use in the electromagnetic environment specified below. Therefore, the customer or
the device user should ensure that it is used in such an environment.
IEC 60601 Test
Immunity Test
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
© 2022 by NeoLight LLC.
All rights reserved.
Electromagnetic Emissions
Compliance
Skylife™ uses R.F. energy only for its internal function.
Group 1
Therefore, its R.F. emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
Class B
Skylife™ is suitable for use in all establishments, including
Class A
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Complies
Electromagnetic Immunity
Compliance Level
± 8kV contact
The relative humidity should be at least 5 %
± 15kV air
± 2 kV for power
The mains power quality should be that of a typical home
supply lines
commercial or hospital environment.
± 1 kV differential
The mains power quality should be that of a typical home
mode
commercial or hospital environment.
The mains power quality should be that of a typical home
0% .5 Periods
commercial or hospital environment. If the user of Skylife™
0% 1 Period
requires continued operation during power mains interruptions,
70% 25 Periods
it is recommended that Skylife™ is powered from an
0% 5 sec
uninterruptible power supply.
Power frequency magnetic fields from common appliances in
the home are not expected to affect the device.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
30A/m
hospital environment. Therefore, keep Skylife™ away from
sources of high levels of power line magnetic fields (above 30
A/m) to reduce the likelihood of interference."
Compliance Level
Level
3 Vrms
6 Vrms in ISM and
amateur radio bands
10 V/m
Electromagnetic Environment – Guidance
Electromagnetic Environment – Guidance
Electromagnetic Environment – Guidance
Skylife™ is suitable for the electromagnetic
environment of typical homes, commercial,
or hospital settings.
Page 40 of 43
500050 Rev A