NeoLight Skylife Service Manual page 27

Infant phototherapy system
Hide thumbs Also See for Skylife:
Table of Contents

Advertisement

CAUTION:
Only use the NeoLight issued light module replacements for the device as any other part
may result in the device being ineffective or dangerous to the user.
CAUTION:
Ensure that the device is turned off and unplugged before proceeding.
1. Turn the Light Bed upside
down. Remove the four
screws next to the
straight rubber feet,
holding the light modules.
Do not unscrew the
screws shown with the
red x marks.
NOTE
2. Carefully remove the old
Light Module from the
Light Bed and rest the
modules on the Light Bed.
Do not strain the
connecting wires by
placing the Light Modules
on the surface you're
working on.
© 2022 by NeoLight LLC.
All rights reserved.
If only one Light Module needs to be replaced, they can be
differentiated before removing the screws. Light Module 1 (a) can be
accessed with the screws closest to the power and data cable.
Therefore Light Module 2 (b) is housed farther from the cable.
b
a
Page 27 of 43
500050 Rev A