3. Carefully remove the
Light Module data port by
pulling the connector.
DO NOT pull the wires to
disconnect the data port
as they may become
damaged. Instead, use
either tweezers or a small
plier to slowly push the
connector out, alternating
from side to side. Go slow
and do not provide too
much force as you may
damage the board.
4. Connect the new Light
Module data port
securely.
Ensure that the four pins
are closest to "J1" for
proper connection. Next,
push on the edges of the
connector with either the
same tweezers, flathead
screwdriver, or very
carefully with the tips of
your fingers.
Make sure to not press on
the wires down during
this process.
5. Gently place the Light module back into the Light Bed. Tighten the screws to 40 in-oz by using a
calibrated torque screwdriver.
IMPORTANT!
Do not over-torque the light module screws as they may damage the device.

4.6 Logging the Light Module Use-life

It is advised to document the Light Module Hours in the provided Service Guide whenever a maintenance
check occurs. To check the hours:
1. Ensure that the Skylife system has been powered on as per operating instructions.
2. Once the device is on, begin a standard treatment by pressing the "+" button. Then pause the
treatment by pressing the power button once to prevent eye strain from blue light.