Chapter 3: Components and Controls ........................12 3.1 System Components: ............................13 Chapter 4: Routine Maintenance Procedures ......................18 4.1 Assembling and Disassembling Skylife ......................18 4.2 Cleaning and Disinfection Procedure (for both professional and home use): ......... 20 4.3 Performance Verification ........................... 22 4.4 Replacing the GelMat............................
The components of the Skylife System must be replaced as directed for the proper function of the system. Please complete service checks listed in Appendix A every half a year (6 months) to ensure all components of the Skylife device are correctly functioning, and replacements are made accordingly.
Skylife is intended for the treatment of neonatal unconjugated hyperbilirubinemia. It is designed to provide phototherapy treatment from underneath the baby. Skylife must be used within a patient enclosure, such as a bassinet, an open crib, a warming table, or an incubator. Skylife can be used in a clinical setting or in the home.
WARNING: Use the Skylife Phototherapy System only for its intended use as described in this manual WARNING: Never operate the system if it has a damaged plug, damaged or frayed power cord, or wires.
Page 10
Monitor the baby's bilirubin levels during phototherapy according to your practitioner's recommendations. WARNING: The Skylife device should be turned off before evaluating the baby's skin color, as lighting will affect visual color evaluation. Parents should contact their medical practitioner if needed.
Page 11
CAUTION: Do not drop the device. If the system is dropped, contact your hospital technician or the manufacturer for further information before using the Skylife unit. CAUTION: Ensure that the device is inaccessible to children or pets when not in use.
Page 17
13. Skylife Carrying Case High-density plastic case with foam inserts to carry Skylife conveniently for administering Home Phototherapy 14. Skylife Travel Case High-density plastic case with foam inserts, rollers, and telescopic handles to carry Skylife safely and conveniently for administering Home Phototherapy.
Chapter 4: Routine Maintenance Procedures This chapter includes procedures for routine maintenance of the Skylife system. These procedures may be performed after any of the following events and/or as prescribed by the institution's maintenance schedule: • Initial receipt of the Skylife at the institution.
Ensure that the disposables do not cover the vents 4. Connect the power cord to the electrical outlet 5. Using the Controller, ensure that Skylife can be turned on/off, the light intensity can be increased/decreased, and the treatment can be paused/un-paused.
Page 21
GelMat may require cleaning. Do not remove the GelMat during cleaning. CAUTION: Do not immerse the Skylife or its accessories in liquid or clean with caustic or abrasive cleaners. Likewise, do not spray or pour any liquid on the Skylife enclosure.
Page 23
2. Apply a disposable swaddle over the Skylife Light Bed as described in section 4.1. Using a pencil, draw a twenty-six cell grid as shown. Each cell is 1" x 1". Ensure that the 7 cell in both rows is located at the center as shown.
8. Spectral measurements are prone to variation in measurement techniques. If the device fails to meet the requirements a 2 time, contact NeoLight customer support. 4.4 Replacing the GelMat CAUTION: Do not clean the Light Bed top surface with any liquids. Instead, use dry cloth wipes only on the cloudy area of the Light Bed.
Page 27
CAUTION: Only use the NeoLight issued light module replacements for the device as any other part may result in the device being ineffective or dangerous to the user. CAUTION: Ensure that the device is turned off and unplugged before proceeding.
It is advised to document the Light Module Hours in the provided Service Guide whenever a maintenance check occurs. To check the hours: 1. Ensure that the Skylife system has been powered on as per operating instructions. 2. Once the device is on, begin a standard treatment by pressing the "+" button. Then pause the treatment by pressing the power button once to prevent eye strain from blue light.
Page 29
If any errors arise during this process, proceed to **(Troubleshooting Section) * for further instructions. 4.7 Packaging Instructions for Skylife Carrying and Travel Cases NOTE This procedure applies to both the Skylife Carrying Case (SKY-CAS) and Skylife travel Case (SKY-TRV) 1. Open out the case on a sufficiently large work surface.
4.8 Homecare Providers Ensure when you deliver Skylife for use in the home that you provide adequate in-service training for parents or guardians. Be sure to provide all the Instructions for Use, Quick Reference Guides, Skylife Disposable Swaddles, and eye covers.
Pressing any key will mute the Warning for 180 seconds " REPLACE!" module is exceeded the Turn-off device recommended limit of Contact NeoLight to order replacement light modules 25,000 hours Follow the "Replacing Light Modules" procedure in Chapter 4.5 to replace the old light modules. - Warning Message:...
Page 34
Likely Cause Trouble Shooting Steps Functionality Issues - Warning Message: Bank 2 EEPROM error Restart the system "!! SYSTEM FAILURE If the error persists, contact NeoLight 00064" - Warning Message: Loose Controller J1 Contact NeoLight "!! SYSTEM FAILURE Connector 00120"...
Page 35
Damaged Parts Cables or Connectors, Incorrect method Remove the unit from service and contact NeoLight or or excessive force trained service personnel. used on All damaged parts must be replaced and the unit checked to attachment and ensure no internal damage has occurred.
Chapter 6: Accessories and Replacement Parts Please contact NeoLight customer Service for any replacements and parts needed. The following parts are critical components to the efficacy and safety of our products and must be replaced as directed. Part Name Catalog Number...
15% and 90% R.H. Stabilization Time The Skylife Phototherapy System does not require a pre-aging time. It is ready for use immediately after turning on the initial time. The Skylife Phototherapy System does not require a stabilization period. It is...
Portable R.F. communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Skylife device, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Page 40
Electromagnetic Emissions Skylife™ is intended for use in the electromagnetic environment specified below. Therefore, the customer or the user of Skylife™ should ensure that it is used in such an environment. Emission Tests Compliance Electromagnetic Environment – Guidance Skylife™ uses R.F. energy only for its internal function.
Product, excluding the Disposable Parts (collectively with the Light Modules, the "Covered Components"). The GelMat, the Skylife Disposable Swaddles, or other expendable or disposable parts of the Product are the "Disposable Parts." NeoLight will replace your Skylife device free of charge, provided the system: •...