7. Ensure that the edges of
the GelMat align to the
edge of the Light Bed and
that the corners of the
GelMat match the blue
line indicators.
8. Starting from the center
on one of the long edges,
lightly press the GelMat
pushing outward to
remove any bubbles,
continue with the same
motion on the other side
and then move until all
edges have adhered.
9. Firmly press the GelMat onto the Light Bed to finish.
10. Perform a final check by running a finger against the edge of the GelMat to ensure it has securely
adhered.

4.5 Replacing Light Modules

1. Nose pliers
2. 40 in-oz Torque Driver with Torx 10 Tip
3. Tweezers
4. Spare screws - 18-8 Stainless Steel Low-Profile Socket Head Screw with Hex Drive, 6-32 Thread
Size, 3/8" Long