Damaged Parts
Cables or Connectors,
Discolored or stained
Diffuser
Tacky GelMat surface
Plastic Enclosure or
Diffuser damaged is
cracked
Out of Specification
Light Output
measurement is out of
specification
© 2022 by NeoLight LLC.
All rights reserved.
1.
Incorrect method
1.
or excessive force
used on
2.
attachment and
removal of the
Controller and
Light Bed besides
shipping purposes
2.
Excess force used
when handling
cables and device
1.
Fluid spilled on the
1.
Diffuser during
GelMat
replacement
1.
Excessive bending
1.
of GelMat can
cause delamination
and force the
encased gel to leak
out through the
seams
2.
A sharp object may
have punctured the
GelMat exposing
the interior gel
1.
The unit was
1.
dropped or
sustained damage
2.
through impact
with another
object.
1.
Improper
1.
measurement
setup.
2.
2.
LED module needs
replacing.
3.
3.
Aged or worn out
GelMat
4.
5.
Remove the unit from service and contact NeoLight or
trained service personnel.
All damaged parts must be replaced and the unit checked to
ensure no internal damage has occurred.
Remove the unit from service and contact NeoLight-
authorized and trained service personnel.
Replace GelMat
Remove the unit from service and contact NeoLight or
trained service personnel.
All damaged parts must be replaced and the unit checked to
ensure no internal damage has occurred.
Ensure that the GelMat and one of the disposables are
appropriately applied
Check if the GelMat has been used for more than 1,500 hrs.
If, so please replace GelMat.
Make sure to use Ohmeda Medical BiliBlanket Light Meter ii
or the meters listed in the conversion table to check light
intensity.
Repeat the Performance Verification steps to confirm. Also,
make sure to hold the meter parallel to the surface and ever
so slightly touch the surface of the disposable while taking
measurements
If the light output is still out of specification, remove the
unit from service and contact NeoLight or trained service
personnel.