7. Compare the adjusted mean values against the specification.
Treatment Setting
Low Setting
High Setting
Very High Setting
8. Spectral measurements are prone to variation in measurement techniques. If the device fails to meet
the requirements a 2

4.4 Replacing the GelMat

CAUTION:
Do not clean the Light Bed top surface with any liquids. Instead, use dry cloth wipes only
on the cloudy area of the Light Bed.
1. 70% Iso Propyl Alcohol
2. Microfiber cloth
NOTE
GelMat should only be replaced by a Biomed technician or other medical professional.
Please document this replacement in the Service Guide provided in Appendices A & B.
1. Remove the old GelMat and dispose of it per institutional practices. The GelMat is made from
cannot
Polyurethane and
2. Clean the Light Bed
plastic surface with 70%
Isopropyl Alcohol and a
microfiber towel to
remove any remaining
adhesive residue. Ensure
that there is no leftover
adhesive or sticky residue
before continuing
© 2022 by NeoLight LLC.
All rights reserved.
��
��������_��������������
Performance Specification
25.0 < I
35.0 < I
55.0 < I
nd
time, contact NeoLight customer support.
Tools and Materials Required
be recycled.
+ δ
= ��
��������
≤ 35.0
mean_adjusted_low
≤ 55.0
mean_adjusted_high
≤ 72.4
mean_adjusted_vhigh
Acceptance Criteria
Pass
Pass
Pass
Page 24 of 43
500050 Rev A