Cleaning And Disinfection Procedure (For Both Professional And Home Use) - NeoLight Skylife Service Manual

Infant phototherapy system
Hide thumbs Also See for Skylife:
Table of Contents

Advertisement

Place the disposable on the
device. Hold the ends of a
longer edge and slide the
distal end beneath the
device as shown. Using the
Skylife to secure the distal
end, pull the proximal end
and slide it beneath the
device.
NOTE:
Ensure that the disposables do not cover the vents
4. Connect the power cord to the electrical outlet
5. Using the Controller, ensure that Skylife can be turned on/off, the light intensity can be
increased/decreased, and the treatment can be paused/un-paused.
WARNING!
During this procedure, ensure that the Controller beeps for each button press. Contact
NeoLight if the Controller does not provide this audible feedback.

4.2 Cleaning and Disinfection Procedure (for both professional and home use):

It is recommended to use Super Sani-Cloth®; however, similar quaternary ammonium disinfectants can be
used provided they have similar Isopropyl Alcohol concentration around 55%, Quaternary ammonium
compounds concentration of 0.25%, and n-Alkyl Dimethyl Benzyl Ammonium Chloride concentration
around 0.25%.
CAUTION:
Unplug the device. Always turn off and disconnect the device from the power outlet
before cleaning.
CAUTION:
Skylife consists of the Lightbed with a controller, GelMat, Skylife Disposable Swaddles, and
the Power Supply. Only the Light Bed, GelMat, and Power Supply are designed to be
reused.
© 2022 by NeoLight LLC.
All rights reserved.
Page 20 of 43
500050 Rev A