Rf Tx/Rx 1 ➀ - Keysight Technologies S9101A Startup Manual

5g multi-band vector transceiver
Hide thumbs Also See for S9101A:
Table of Contents

Advertisement

2. Calibrate the Cable signal path
3. Calibrate the Receiver (Rx) signal path
S9101A 5G Multi-Band Vector Transceiver, Startup Guide
a. Disconnect the power sensor from the
S9101RH mmWave Transceiver 1 connector
b. Using an 8121-3222 cable, 2.4 mm (m) to 2.4 mm (m), 152.4 mm (6.0 in)
or equivalent, connect one end of the cable to the
S9101RH mmWave Transceiver 1 connector
c. Using an 11900B adapter, 2.4 mm (f) to 2.4 mm (f) or equivalent,
connect the U8487A Power Sensor or equivalent to the end of the cable.
d. Select
to continue the calibration.
Ok
View the
Results Cable
a. Disconnect the power sensor and adapter from the cable.
b. Connect the free end of the cable to the
S9101RH mmWave Transceiver 1 connector
c. Select
to complete the calibration.
Ok
View the
Results Receiver
Corrections are stored in:
Rx-RFInputOutput2-CHANNELPOWER.csv
RF Tx/Rx 1 ➀
RF Tx/Rx 1 ➀
tab when the calibration is complete.
RF Tx/Rx 2 ➁
tab when the calibration is complete.
S9100ID_<SerialNumber>-
7 Run Power Calibration
.
.
.
83