Calibrating S9101A-Tr2 - Keysight Technologies S9101A Startup Manual

5g multi-band vector transceiver
Hide thumbs Also See for S9101A:
Table of Contents

Advertisement

7. 4. 3. 3 Calibrating S9101A-TR2

Calibration 1 of 4: Power from RRH 1 RFHD 1 -> RRH 1 RFHD 2, Transceiver 1
1. Calibrate the Source (Tx) signal path
S9101A 5G Multi-Band Vector Transceiver, Startup Guide
This calibration is used to calibrate the FR2 mmWave ports of
mmWave Transceiver 1 (with
RF Output) on an S9101A-TR2 5G Multi-Band Vector Transceiver.
When performing this Power Calibration, the Source (Tx) output
power level is set to -5 dBm by default.
a. Select the Power Calibration tab.
b. Select the Operation drop-down arrow and select
c. Select the Calibration Step(s) drop-down arrow and select
7 Run Power Calibration
RF Tx/Rx 1 ➀
used as the
Calibration
.
.
Complete
81