Waters 2998 Operator's Manual page 50

Photodiode array detector
Hide thumbs Also See for 2998:
Table of Contents

Advertisement

5.
Reinstall the detector front left panel cover:
To prevent invalid subsequent chromatographic results, remove the
cuvette from the detector, and replace the empty holder after
running your cuvette measurements.
To maintain optimum system performance, replace the front left
panel cover before resuming normal operation of the detector.
6.
Insert a reference cuvette containing the mobile phase standard, and
run a zero measurement.
7.
Replace the reference cuvette with a cuvette containing your analyte
dissolved in mobile phase solvent, and run a sample measurement.
8.
Use the storage, review, subtract and review, and replay functions to
analyze the data obtained.
2-24
Setting Up the Detector