Waters SAT/IN™ Module
Waters, or other, manual injector
Waters autopurification system
Other manufacturers' integrator or A/D interface device
To avoid electrical shock, power-off instruments being
Warning:
connected before making any electrical connections.
To meet the regulatory requirements of immunity from
Requirement:
external electrical disturbances that can affect the performance of this
instrument, do not use cables longer than 3 meters (9.8 feet) when you make
connections to the I/O connectors. In addition, ensure you connect the shield of
the cable to ground at one instrument only.
To connect the detector to other instruments, use the two analog-out/event-in
(I/O) connectors and their mating connectors on the rear panel.
2998 rear panel analog-out/event-in connectors
Connector B (outputs)
Connector A (inputs)
+
1
Analog 1
2
Analog 1
3
Ground
+
4
Analog 2
5
Analog 2
+
6
Switch 1
7
Switch 1
8
Ground
+
9
Switch 2
10
Switch 2
+
1
Inject Start
2
Inject Start
3
Ground
+
4
Lamp Off
5
Lamp Off
+
6
Chart Mark
7
Chart Mark
8
Ground
+
9
Auto Zero
10
Auto Zero
Making signal connections