Waters 2998 Operator's Manual page 116

Photodiode array detector
Hide thumbs Also See for 2998:
Table of Contents

Advertisement

Mobile phase absorbance measured against air or water (Continued)
Absorbance at specified wavelength (nm)
200
Acids and bases
Acetic acid,
2.61
1%
Hydro-
0.11
chloric acid,
0.1%
Phosphoric
<0.01 —
acid, 0.1%
Trifluoro-
1.20
acetic acid
Diammo-
1.85
nium
phosphate,
50 mM
Triethy-
2.33
lamine, 1%
Buffers and Salts
Ammonium
1.88
acetate,
10 mM
Ammonium
0.41
bicarbonate,
10 mM
EDTA,
0.11
disodium,
1 mM
HEPES,
2.45
10 mM,
pH 7.6
MES,
2.42
10 mM,
pH 6.0
C-16 Solvent Considerations
205
210
215
2.63
2.61
2.43
0.02
<0.01 —
0.78
0.54
0.34
0.67
0.15
0.02
2.42
2.50
2.45
0.94
0.53
0.29
0.10
0.01
<0.01 —
0.07
0.06
0.04
2.50
2.37
2.08
2.38
1.89
0.90
220
230
240
2.17
0.87
0.14
0.22
0.06
<0.02 <0.01 —
<0.01 —
2.37
1.96
0.50
0.15
0.02
<0.01 —
0.03
0.03
0.02
1.50
0.29
0.03
0.45
0.06
<0.01 —
250
260
280
0.01
<0.01 —
0.12
0.04
<0.01
0.02
0.02
0.02
<0.01 —