Alarm settings range and default values
Setting
Low O
2
High O
)
2
Low VE
High VE
Low VTE
High VTE
Circuit leak
Cardiac bypass
1009-0633-000
4 Alarms and Troubleshooting
Range
Increment
18-99%
1%
21-99%, Off
1%
Off, 0.1-10 L/min
0.1 L/min
0.5-30, off L/min
0.5 L/min
Off, 5-1500 mL
5 mL if <20 mL,
20 otherwise
20 - 1600 mL, off
20 mL
Audio on or Audio off
n/a
No or In Progress
n/a
Default
21%
Off
2.0 L
10.0 L
Off
1000 mL
Audio on
No
4-17