Electromagnetic compatibility

w WARNING
6-8
Changes or modifications to this equipment not expressly approved by Datex-
Ohmeda could cause EMC issues with this or other equipment. Contact Datex-
Ohmeda for assistance. This device is designed and tested to comply with
applicable regulations regarding EMC as follows.
Use of portable phones or other radio frequency (RF) emitting
equipment near the system may cause unexpected or adverse
operation. Monitor operation when RF emitters are in the vicinity.
• Use of other electrical equipment on or near this system may
cause interference. Verify normal operation of equipment in
your configuration before use on patients.
The system provides connections for items such as printers, visual displays
and hospital information networks. When these items (non-medical
equipment) are combined with the S/5 Aespire, these precautions must be
followed:
• Don't place items not approved to IEC 60601-1 closer than 1.5 m to the
patient.
• All items (medical electrical equipment or non-medical electrical
equipment) connected to the S/5 Aespire by a signal input/signal output
cable must be supplied from an AC power source which uses a separating
transformer (in accordance with IEC 60989) or be provided with an
additional protective earth conductor.
• If a portable multiple socket outlet assembly is used as an AC power
source, it must comply with IEC 60601-1-1. The assembly must not be
placed on the floor. Using more than one portable multiple socket outlet
assembly is not recommended.
Do not connect non-medical electrical equipment directly to the AC outlet at
the wall instead of an AC power source which uses a separating transformer.
Doing so may increase enclosure leakage current above levels allowed by IEC
60601-1 in normal conditions and under single-fault conditions. This may
cause an unsafe electrical shock to the patient or operator.
The S/5 Aespire provides multiple AC outlet sockets for connecting only other
medical electrical equipment. Do not connect non-medical electrical
equipment to these sockets. Doing so may increase enclosure leakage
current above levels allowed by IEC 60601-1 in normal conditions and under
single-fault conditions. This may cause an unsafe electrical shock to the
patient or operator.
After connecting anything to these outlets, conduct a complete system
leakage current test (according to IEC 60601-1).