Ventilator Accuracy Data; Delivery Accuracy; Monitoring Accuracy - Datex-Ohmeda Aespire 7900 User's Reference Manual

Hide thumbs Also See for Aespire 7900:
Table of Contents

Advertisement

Aespire 7900

Ventilator accuracy data

Delivery accuracy

Monitoring accuracy

6-26
The following accuracy data are based on patient conditions and settings
described in ASTM F1101. The ventilator is assumed to be operating in
volume mode, with 100 percent oxygen in the breathing system; or, it is
connected to an anesthesia gas analyzer. If the ventilator is operating without
being connected to an anesthesia gas analyzer, additional errors may occur
which are described in the "Gas composition related errors (both modes)"
chart which is provided on the next page.
The minimum detectable breath size is 5.0 mL.
Volume control
>210 mL tidal volume: Better than ±7% of set V
(100% O
)
60 to 210 mL tidal volume: Better than ±15 mL
2
<60 mL tidal volume: Better than ±10 mL
Pressure mode
Inspiratory pressure: ±2.0 cm H
(100% O
)
PEEP: ±1.5 cm H
2
greater
Volume control
>210 mL tidal volume: Better than ±9% of V
(100% O
)
60 to 210 mL tidal volume: Better than ±18 mL
2
<60 mL tidal volume: Better than ±10 mL
Pressure mode
Better than: ±2.0 cm H
Measured values are taken at ambient temperature and pressure, and are not
corrected when displayed.
0 (repeatability only)
2
O: greater of ±10% or ±3 cm H
2
T
0 or ±5% of reading whichever is greater
2
T
O, whichever is
2
1009-0633-000

Advertisement

Table of Contents
loading

Table of Contents