Aespire 7900
Immunity test
IEC 60601-1-2 test level
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8 3
A/m
Note: U
is the AC mains voltage before application of the test level.
T
6-10
Compliance level
Electromagnetic environment guidance
3 A/m
If display distortion or other abnormalities occur, it
may be necessary to position the S/5 Aespire
Anaesthetic System further from sources of power
frequency magnetic fields or to install magnetic
shielding. The power frequency magnetic field
should be measured in the intended installation
location to assure that it is sufficiently low.
1009-0633-000