Alarm Setup

Setting volume
alarms
WARNING

Alarm limit setup

M1132382
Use the volume alarms key to turn the volume alarms on and off.
When the alarms are off, an X covers the alarm limits. Use this control
during manual ventilation when constant attention is on the patient.
Use the End case key to minimize alarms between cases. The alarms
will reactivate when two or more breaths are detected within 30
seconds.
Do not turn off volume alarms for a spontaneously
breathing patient. The system will not alarm for low
volume.
To set or change alarm limits:
1. Push the Menu key.
2. Select Alarm Setup from the Main Menu.

Alarm Setup

O2
MV
TVexp
Leak Audio
Go to Main Menu
3. To set an alarm limit, use the ComWheel to scroll to the desired
alarm limit. Push the ComWheel to select the limit. The following
alarm limits can be set or changed:
O2, MV, and TVexp - Alarm limits for both high and low can
be set. The low limit is the left numeric field, and the high limit
is the right numeric field.
Leak Audio - The patient circuit leak alarm activates during
mechanical ventilation if less than half of the inspired volume
returns through the expiratory flow sensor. Prevent expected
alarms from known circuit leaks by setting the Leak Audio to
Off.
4. Change the value with the ComWheel.
5. Select Go to Main Menu to return to the Main Menu, or push the
Menu key to return to the normal monitoring screen.
21
99
3.0
7.0
200
600
On
3 Operation
3-5