Datex-Ohmeda Aespire 7900 User's Reference Manual page 24

Hide thumbs Also See for Aespire 7900:
Table of Contents

Advertisement

Aespire 7900
w WARNING
w CAUTION
2-2
Obey applicable safety precautions:
• Read the material safety data sheet for each cleaning agent.
• Read the operation and maintenance manual for all
sterilization equipment.
• Wear gloves and safety glasses. A damaged O
leak and cause burns (contains potassium hydroxide).
• Do not breathe the fumes.
To help prevent damage:
• Refer to the manufacturer's data if you have questions about a
cleaning agent.
• Do not use organic, halogenated, or petroleum based
solvents, anesthetic agents, glass cleaners, acetone, or other
harsh cleaning agents.
• Do not use abrasive cleaning agents (such as steel wool, silver
polish or cleaner).
• Keep all liquids away from electronic parts.
• Do not permit liquid to go into the equipment housings.
• Do not soak synthetic rubber parts for more than 15 minutes.
Swelling or faster aging can occur.
• Only autoclave parts marked 134°C.
• Cleaning solutions must have a pH of 7.0 to 10.5.
sensor can
2 2 2 2
1009-0633-000

Advertisement

Table of Contents
loading

Table of Contents