Flow sensor zeroing

wWARNING
Step 1
Pull the latch (1) to unlock the
flow sensor module from the
breathing system.
Step 2
Pull the flow sensor module from
the breathing system.
Step 3
When zeroing is complete, the
screen shows, "No Insp flow
sensor" and "No Exp flow
sensor."
Step 4
Install the flow sensor module.
Step 5
Do a breathing circuit leak test before using the system. Refer to "Breathing system tests" in section 4 Preoperative Tests of the
User Reference Manual, Part 1.
1009-0633-000
Do not perform calibration while unit is connected to a patient.
The system automatically corrects for zero offset when you unplug the flow
sensor connectors with power on. You must stop mechanical ventilation
before you calibrate the flow sensors.
Note: Properly performing the O
1
N N N N o o o o I I I I n n n n s s s s p p p p F F F F l l l l o o o o w w w w S S S S e e e e n n n n s s s s o o o o r r r r
N N N N o o o o E E E E x x x x p p p p F F F F l l l l o o o o w w w w S S S S e e e e n n n n s s s s o o o o r r r r
3 User Maintenance
cell zeroing will also zero the flow sensor.