Datex-Ohmeda Aespire 7900 User's Reference Manual page 113

Hide thumbs Also See for Aespire 7900:
Table of Contents

Advertisement

Gas composition related
1009-0633-000
errors
20
30
0
10
20
30
40
50
60
> -10%
70
80
Note
Gas composition errors may be in addition to the above normalized accuracy.
When adding errors, positive errors can have the effect of nulling out negative
errors.
Note
Use of anesthetic agent could affect the errors by approximately -0.95%/%
volume agent.
When subjected to gas mixtures containing the following concentrations of
gases, the oxygen monitor has been tested to be within ±5% of the actual gas
concentration. Gas mixtures other than the ones listed below may result in an
accuracy of the oxygen monitor outside of the ±5% V/V.
Gas:
Carbon dioxide
Nitrous oxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
6 Specifications and Theory of Operation
% O
2
40
50
60
70
> +10%
> ± 5%
At concentration:
5%
80%
4%
5%
5%
5%
15%
80
90
100
6-27