Datex-Ohmeda Aespire 7900 User's Reference Manual page 58

Hide thumbs Also See for Aespire 7900:
Table of Contents

Advertisement

Aespire 7900
The screen shows 'Calibrating ... ', followed by the result: 'Complete' or 'Failure'.
If the calibration fails,
• do it again or
• decrease the airway pressure. Then try again.
After repeated failures replace the O
Continue with steps 14 through 16 to exit to a normal screen.
Step 15
To exit, select "Go to Cal
Menu" and push the knob.
You can also press
the menu key to exit to
the normal screen
Step 16
Select "Go to Main Menu"
and push the knob.
Step 17
Select "Exit to Normal
Screen" and push the knob.
Step 18
Do a breathing circuit leak test before using the system.
3-10
sensor and re-calibrate at 21%.
2
Go to Cal Menu
Go to Main Menu
Exit to Normal Screen
1009-0633-000