Aespire 7900
2-10
8. On the manifold assembly, remove the check valves circuit lens (1) by
squeezing the latches (2) together and pulling up on the lens.
1
2
9. Press at (2) to unlock the ramp (1). Rotate the ramp and remove the tabs
from the slots (3) to remove the ramp.
10. Lift the diaphragm to remove it.
1
2
3