100% O
sensor
2
calibration
w WARNING
Step 11
Turn then push the knob to
select O
Sensor Cal.
2
Step 12
Select 100% then push the
knob.
Step 13
With the O
sensor in the
2
circuit, fill the circuit with
100% O
:
2
• Push the flush
button.
•
flow 100% O
Then
5 L/min. (Circuit should
be open.)
Step 14
Select Start Cal. Then push
the knob to start calibration.
1009-0633-000
This procedure takes three minutes or less.
You must complete the 21% calibration before you can select the 100%
calibration.
Do not perform calibration while unit is connected to a patient.
at
2
3 User Maintenance
3-9