Cleaning, maintenance and dispatch

8.2 Servicing

Independently of the use of the device, we recommend regular visual inspection /
servicing of the PRIMEDIC™ HeartSave PAD and the accessories by the user / service
technician at least once a year.
Make sure that the housing, cable, SavePads and all the other accessories are
undamaged.

8.2.1 Servicing check list

When doing so, pay attention to the following:
Note
Operating instructions
38
Check the expiry date
of the SavePads
of the battery pack (optional) and
if necessary replace the parts.
Check whether
the status display "OK" is showing.
you can switch on the device.
the device automatically carries out the self-test after being switched on.
the slot for the power supply is clean.
the device is fully equipped.
If parts of the housing or insulation are damaged, they
must be repaired or exchanged immediately.
If parts of the housing or insulation are damaged, do not
use the device or switch it off immediately.
Have the device repaired as soon as possible by the
manufacturer.
For more detailed information on the regular safety and metrological checks
in accordance with the Ordinance on the Operation and Use of Medical
Devices (MPBetreibV), refer to the Appendix.