Repair and Replacement Procedures
Removing and Replacing the A60–A63 70 GHz Doublers
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests
Figure 7-31
A60 through A63 70 GHz Doublers Removal
1. The A26 splitter (5067-4086) and N5247-20111, N5245-20022, N5245-20023, and N5247-20110 cables are only used
with a legacy HMA26.5 p/n: 5087-7765. The N5240-20125, N5247-20184 and N5247-20185 cables are used only with the
new N5240-60101 HMA-26.5. If you are unclear which HMA26.5 assembly your PNA has installed, refer to Chap-
ter 7 Repairs and
Figure 7-17 on page 38
option-model in Chapter 6
Configuration, Serial Number Prefixes <6021" on page
Keysight N5247B Service Guide
Torque all cable connections to 10 in-lbs.
that pertain to this removal procedure. Refer to
and for details on A26 splitter and cabling, refer to your
"2-Port Configurations, Serial Number Prefixes <6021" on page 6-31
Table 7-2 on page
1
6-84.
7-92.
and
"4-Port
7-63