Troubleshooting
Measurement System Troubleshooting
5. If the A15 DDS assembly is faulty, contact Keysight Technologies for help.
Figure 4-10
Example of Power On Self Test Error
4-38
c. Verify that all cable connections to the A15 DDS assembly are
connected properly.
d. Repeat steps 1 through.3.
e. If the instrument still has a similar "Slug" failure then the DDS is
faulty and needs to be replaced.
Refer to
"Contacting Keysight" on page
2-7.
Keysight N5247B Service Guide